If your UrSpring unit reports an error, the red LED on the controller will blink.
To view the error codes:
1. Go to menu View alerts.
2. The display shows Last Error ------ .
3. Use and to browse through the errors.
4. To confirm an error, press the enter-button.
Error Code |
Possible cause |
Solution |
101 |
The emergency flush is running or has run. A flush is activated for 20 seconds. Because the maximum transmembrane pressure (feed to filtrate) is exceeded (> 2.5 bar/36psi). |
Seek and eliminate the cause of the high transmembrane or absolute pressure. It can damage the UrSpring
|
102 |
Permeability of the filter module is too low. |
Contact your service technician for further help |
103 |
Filter module should be changed soon. |
Prepare the replacement of the filter module. |
104 |
The last flush cleaning did not raise the performance of the filter above the set threshold value for cleaning. |
At the Flush if fouled setting (Automatic flush when fouled):
The UrSpring should now incrementally regain its performance in the course of the next few days. If the performance is not increased after a few days, change the filter module and consider a more regular flushing (as described in the steps above). |
105 |
Problem with power supply. |
Check the power supply and socket. |
106 |
Ambient temperature above 55°C (131°F). |
Lower the ambient temperature. |
107 |
Ambient temperature below 0°C (32°F). |
Make sure the water does not freeze. The electronic control will work below 0°C. |
109 |
Pressure sensors are not functioning correctly. |
Calibrate pressure sensor (see Calibrating the pressure sensors) If the error remains, replace the sensors (Exchanging the pressure sensors) or contact your service technician. |
111 112 113 |
Time setting was lost or could not be saved. |
Make sure the power supply works and reset the time (Set the correct time). |
115 |
Memory error. |
It might be that the power supply of the UrSpring was lost for a longer time, make sure the power supply and socket work correctly. Contact your service technician. |
For further support, please contact Support through our Service Desk. You find the required contact data under https://support.seccua.com/hc/en-us.